Perioperative non-invasive versus semi-invasive cardiac index monitoring in patients with bariatric surgery - a prospective observational study
Results
Sixty patients (mean BMI 49.2 kg/m2) were enrolled into the study and data from 54 patients could be entered in the final analysis. Baseline CI was 3.2 ± 0.9 and 3.3 ± 0.8 l/min/m2, respectively. Pooled absolute CI values showed a positive correlation (rs = 0.76, P < 0.001) and mean bias of of - 0.16 l/min/m2 (limits of agreement: - 1.48 to 1.15 l/min/m2) between the two methods. Pooled percentage error was 56.51%, missing the criteria of interchangeability (< 30%). Preoperatively, bias ranged from - 0.33 to 0.08 l/min/m2 with wide limits of agreement. Correlation of CI was best (rs = 0.82, P < 0.001) and percentage error lowest (46.34%) during anesthesia and after fluid bolus administration. Intraoperatively, bias ranged from - 0.34 to - 0.03 l/min/m2 with wide limits of agreement. CI measurements correlated best during pneumoperitoneum and after fluid bolus administration (rs = 0.77, P < 0.001; percentage error 35.95%). Trending ability for all 10 measurement points showed a concordance rate of 85.12%, not reaching the predefined Critchley criterion (> 92%).
Conclusion
Non-invasive as compared to semi-invasive CI measurements did not reach criteria of interchangeability for monitoring absolute and trending values of CI in morbidly obese patients undergoing bariatric surgery.
Trial registration
The study was registered retrospectively on June 12, 2017 with the registration number NCT03184272 .