Comparison of oscillometric, non-invasive and invasive arterial pressure monitoring in patients undergoing laparoscopic bariatric surgery - a secondary analysis of a prospective observational study

Results
Sixty patients (mean body mass index of 49.2 kg/m2) were enrolled and data from 56 finally analysed. Pooled blood pressure values of all time points showed a significant positive correlation for both NIPB and Nexfin® versus IAP. Pooled PE for NIBP versus IAP was 37% (SAP), 35% (DAP) and 30% (MAP), for Nexfin versus IAP 23% (SAP), 26% (DAP) and 22% (MAP). Correlation of MAP was best and PE lowest before induction of anesthesia for NIBP versus IAP (r = 0.72; PE 24%) and after intraoperative fluid bolus administration for Nexfin® versus IAP (r = 0.88; PE: 17.2%). Concordance of MAP trending was 90% (SAP 85%, DAP 89%) for NIBP and 91% (SAP 90%, DAP 86%) for Nexfin®. MAP trending was best during intraoperative ATP positioning for NIBP (97%) and at induction of anesthesia for Nexfin® (97%).

Conclusion

As compared with IAP, interchangeability of absolute pressure values could neither be shown for NIBP nor Nexfin®, however, NIBP showed poorer overall correlation and precision. Overall trending ability was generally high with Nexfin® surpassing NIBP. Nexfin® may likely render individualized decision-making in the management of different hemodynamic stresses during laparoscopic bariatric surgery, particularly where NIBP cannot be reliably established.

Trial registration

The non-interventional, observational study was registered retrospectively at ( NCT03184285 ) on June 12, 2017.

Rights

Use and reproduction:


CC BY 4.0

Please note that individual components of the publication may be subject to other licensing or copyright conditions.

Cite

Citation style:
Could not load citation form.