Magnetoneurograhy of an Electrically Stimulated Arm Nerve

For further development and optimization of novel uncooled  magnetoelectric  (ME)  sensors,  a  better  understanding of spectral power distribution and signal strength of nerve signals is of high interest. For obtaining information on these  signal  properties,  Superconducting  Quantum  Interference Device (SQUID) measurements were performed at the Physikalisch-Technische  Bundesanstalt  (PTB)  in  Berlin. Signal  amplitudes  were  subject-dependent  and  ranged  from 
17 fT to 60 fT in a frequency range from 100 Hz to 1 kHz. The  required  SQUID  averaging  time  was  in  the  range  of minutes,  while  for  current  ME  sensors  significantly longer averaging times are expected to be necessary. 

Rechte

Nutzung und Vervielfältigung:


CC BY-NC-ND 4.0

Bitte beachten Sie, dass einzelne Bestandteile der Publikation anderweitigen Lizenz- bzw. urheberrechtlichen Bedingungen unterliegen können.

Zitieren

Zitierform:
Zitierform konnte nicht geladen werden.